JHI-08
JHI-08
World First Hybrid Pyrolyzer
World First Hybrid Pyrolyzer
Maximize the benefit of Hybrid
Two heating methods by "Curie Point" and "Furnace" are both available in one system. It is the most versatile pyrolyzer that offers 3 different analysis modes.
Pyrolysis Mode
Two ways of pyrolysis are available. You can choose Curie Point pyrolysis using Pyrofoils or Furnace pyrolysis with ceramic heater depending on sample shapes and target.
Hybrid Mode
You can conduct thermal extraction and Curie Point pyrolysis on one sample continuously without interruption.
EGA Mode (Evolved Gas Analysis)
You can program multiple phases of temperature elevation with different duration and rate. It is usually directly coupled with a mass spectrometer to find thermal decomposition behavior of samples.
Easily removable module prevents cross contamination in the flow chancel
During pyrolysis, approximately 60% of tar-like high-boiling materials remain in the sample tube. This is especially a problem for samples of molded resins and rubbers that contain inorganic filler, carbon powder and other additives. JHI-08 has easily replaceable sample tube and needle assy for cleaning . This minimizes cross contamination and results in better and reproducible chromatograms.
Maximize the benefit of Hybrid
Two heating methods by "Curie Point" and "Furnace" are both available in one system. It is the most versatile pyrolyzer that offers 3 different analysis modes.
Pyrolysis Mode
Two ways of pyrolysis are available. You can choose Curie Point pyrolysis using Pyrofoils or Furnace pyrolysis with ceramic heater depending on sample shapes and target.
Hybrid Mode
You can conduct thermal extraction and Curie Point pyrolysis on one sample continuously without interruption.
EGA Mode (Evolved Gas Analysis)
You can program multiple phases of temperature elevation with different duration and rate. It is usually directly coupled with a mass spectrometer to find thermal decomposition behavior of samples.
Curie Point Heating method
Curie Point Pyrolysis using Pyrofoil has an established reputation with over 40 years of history and produces reliable and sharp chromatogram. It reaches up to the target temperature in an instance. The time it takes is as fast as 0.2 seconds to any temperature that unwanted secondary reactions and decomposition will not happen during the waiting time.
Easily removable module prevents cross contamination in the flow chancel
During pyrolysis, approximately 60% of tar-like high-boiling materials remain in the sample tube. This is especially a problem for samples of molded resins and rubbers that contain inorganic filler, carbon powder and other additives. JHI-08 has easily replaceable sample tube and needle assy for cleaning . This minimizes cross contamination and results in better and reproducible chromatograms.
Options to make it more versatile
Auto sampler unit APC-50 (Click here for details)
APC-50 Auto sampler automates the sample feeding/collection process for both Pyrofoils and sample cups. Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APS-50 Auto sampler for Pyrofoils only is also available.
Options to make it more versatile
Auto sampler unit APC-50 (Click here for details)
APC-50 Auto sampler automates the sample feeding/collection process for both Pyrofoils and sample cups. Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APS-50 Auto sampler for Pyrofoils only is also available.
Pyrolysis in air atmosphere and Column inlet cooler
Purge Gas Selector (PGS) allows to switch carrier gas to air easily. By combining with Column Inlet Cooler (CIC), you can perform pyrolysis under ambient air.
The generated gas is split again (2nd splitter) after GC injection port. One goes to GCMS for standard analysis, the other is vent out and collected to mini-PAT (Sample tube with TenaxGR and 280°C Pyrofoil). The trapped gas in mini-PAT can be reanalyzed using JCI-77.
Optional Purge Gas Selector (PGS) allows to switch carrier gas to air easily. By combining with Column Inlet Cooler (CIC), you can perform pyrolysis under ambient air.
The generated gas is split again (2nd splitter) after GC injection port. One goes to GCMS for standard analysis, the other is vent out and collected to mini-PAT (Sample tube with TenaxGR and 280°C Pyrofoil). The trapped gas in mini-PAT can be reanalyzed using optinal JCI-77.
Specifications
Specifications
Curie point pyrolysis (CP-PY) | |
Heating method | Curie point heating |
Sample setting | by dropping a Pyrofoil in the oven |
Pyrolysis temp. | by choosing a Pyrofoil Temp. range from 160°C to 1040°C |
Pyrolysis time | 1 to 99 sec |
Needle | 0.64 mm OD x 70 mm |
Sample tube material | Quartz glass |
Dimentions, Weight | W120 x D121 x H328 mm, 2 kg |
Furnace pyrolysis (F-PY) | |
Heating method | Furnace oven heating |
Heater | Tubular ceramic heater |
Sample setting | by dropping Pyrofoil or sample cup in the oven or lower the holder by hand |
Pyrolysis temp. | 40 to 800°C by 1°C unit |
Temp. elevation rate | Max 600°C / min |
Needle | 0.64 mm OD x 70 mm |
Sample tube material | Quartz glass |
Dimensions, Weight | W120 x D121 x H328 mm, 2 Kg |
Controller unit | |
Needle heater | 40°C - 300°C |
Oven heater | Ceramic heater 40°C - 800°C |
Radio frequency generator | Radio frequency output : 48W Fall detect sensor is installed |
Carrier gas | GC/JHI Switching valve is installed |
Dimensions, Weight | W180 x D500 x H390 mm, 16 Kg |
Power | 100 - 220V, 500VA (Max) |
Curie point pyrolysis (CP-PY) | |
Heating method | Curie point heating |
Pyrolysis time | 1 sec to 999 sec |
Sample setting | by dropping a Pyrofoil in the oven |
Pyrolysis temp. | by choosing a Pyrofoil Temp. range from 160°C to 1040°C |
Furnace pyrolysis (F-PY) | |
Heating method | Furnace oven heating |
Heater | Tubular ceramic heater |
Sample setting | by dropping a Pyrofoil or sample cup in the oven or lower the holder by hand |
Pyrolysis temp. | 40 to 800°C by 1°C unit |
Temp. elevation rate | Max 600°C / min |
Pyrolysis time | 1 to 99 sec |
Needle | 0.64 mm OD x 70 mm |
Sample tube material | Quartz glass |
Dimension, Weight | W120 x D121 x H328 mm, 2 Kg |
Controller unit | |
Needle heater | 40°C - 300°C |
Oven heater | Tubular ceramic heater 40°C - 800°C |
Radio frequency generator | Radio frequency output : 48W Fall detect sensor is installed |
Carrier gas | GC/JHI Switching valve is installed |
Dimensions, Weight |
W180 x D500 x H390 mm, 16 Kg |
Power | 100 - 220V, 500VA (Max) |
Auto Sampler for Hybrid Pyrolyzer JHI-08
Auto sampler for Hybrid Pyrolyzer
Auto Sampler APC-50
APC-50 Auto sampler automates the sample feeding/collection process for both Pyrofoils and sample cups. Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APS-50 Auto sampler for Pyrofoils only is also available.
Two ways of sample feeding
Having two types of turntable, APC-50 supports sample feeding for both Pyrofoils and sample cups. Two turntables come as a standard accessory.
Turntable for Pyrofoils
Turntable for sample cups
Pyrofoil and Cup sampler APC-50 | |
Maximun Number of samples | 50 samples at a time |
Sample introduction | By turntable |
Sample Retrieval | By magnetic retriever stick |
Air Purge | Auto purge by solenoid valve |
Dimensions, Weight | W175 x D265 x H662 mm including JHI-08 Pyrolyzer, 8 kg |
APC-50 installed on JHI-08
Auto Sampler APC-50
APC-50 Auto sampler automates the sample feeding/collection process for both Pyrofoils and sample cups. Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APS-50 Auto sampler for Pyrofoils only is also available.
Two ways of sample feeding
Having two types of turntable, APC-50 supports sample feeding for both Pyrofoils and sample cups. Two turntables come as a standard accessory.
Turntable for Pyrofoils
Turntable for sample cups
Pyrofoil and Cup sampler APC-50 | |
Maximun Number of samples | 50 samples at a time |
Sample introduction | By turntable |
Sample Retrieval | By magnetic retriever stick |
Air Purge | Auto purge by solenoid valve |
Dimensions, Weight | W175 x D265 x H662 mm including JHI-08 Pyrolyzer, 8 kg |
Pyrofoil Sampler
Pyrofoil sampler for Hybrid Pyrolyzer
Auto Sampler APS-50
APS-50 Auto sampler automates the sample feeding/collection process for Pyrofoils (Sample cups are not supported). Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APC-50 Auto sampler for Pyrofoils and sample cups is also available.
Sample feeding
Maximum 50 pyrofoils can set at a time.
Foil magazine
Pyrofoil and Cup sampler APS-50 | |
Maximun Number of samples | 50 samples at a time |
Sample introduction | By turntable |
Sample Retrieval | By magnetic retriever stick |
Air Purge | Auto purge by solenoid valve |
Dimensions, Weight | W275 x D272 x H612 mm including JHI-08 Pyrolyzer, 6 kg |
APS-50 installed on JHI-08
Auto Sampler APS-50
APS-50 Auto sampler automates the sample feeding/collection process for Pyrofoils (Sample cups are not supported). Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APC-50 Auto sampler for Pyrofoils and sample cups is also available.
APS-50 installed on JHI-08
Sample feeding
Maximum 50 pyrofoils can set at a time.
Foil magazine
Pyrofoil and Cup sampler APS-50 | |
Maximun Number of samples | 50 samples at a time |
Sample introduction | By turntable |
Sample Retrieval | By magnetic retriever stick |
Air Purge | Auto purge by solenoid valve |
Dimensions, Weight | W275 x D272 x H612 mm including JHI-08 Pyrolyzer, 6 kg |
Pyrolyzer