FOC-06
FOC-06
Simple Outgas collector
50 mL and 100 mL glass vials are available to heat samples and collect volatilized gas into mini-PAT.
Number of sample
Able to collect VOC from maximum of 6 samples.
Sample Chamber
Always place a sample in clean vial(consumable) to save time from cleaning process.
Thermal extraction temperature
From ambient to maximum of 250°C
Break-through prevented
If the sample contains break-through easy compounds, mini-PAT is able to connect two pieces in series to improve collection.
Other sample tubes
In addition to mini-PAT, customized joint connections are available to fit other sample tubes.
Number of sample
Able to collect VOC from maximum of 6 samples.
Sample chamber
Always place a sample in clean vial (consumable) to save time from cleaning process.
Thermal extraction temperature
From ambient to maximum of 250°C
Break-through prevented
If the sample contains break-through easy compounds, mini-PAT is able to connect two pieces in series to improve collection.
Other sample tubes
In addition to mini-PAT, customized joint connections are available to fit other sample tubes.
Specifications
FOC-06 | |
Number of samples | 6 |
Sample vial material | Glass |
Sample vial volume | 50mL or 100mL |
Heating method | Resistance heating |
Heating temp. range | Ambient to 250°C (300°C max when baking) |
Sampling gas flow | 1000mL/min in total |
Sampling gas | He or N2 |
Dimensions | Oven unit 245(W)×140H)×200(D)mm Controller unit 167(W)×180(H)×210(D)mm |
Weight | Oven unit 5.8Kg Controller unit 4Kg |
Power | AC100V 500VA |
FOC-06 | |
Number of samples | 6 |
Sample vial material | Glass |
Sample vial volume | 50 mL or 100 mL |
Heating method | Resistance heating |
Heating temp. range | Ambient to 250°C (300°C max when baking) |
Sampling gas flow | 1000 mL/min in total |
Sampling gas | He or N2 |
Dimensions | Oven unit 245(W) x 140H) x 200(D) mm Contoroller unit 167(W) x 180(H) x 210(D) mm |
Weight | Oven unit : 5.8 Kg Contoroller unit : 4 Kg |
Power | AC100V 500VA |