Home | Products | Curie Point Injector | JPI-15Ⅱ_JPI-07Ⅱ

JPI-15Ⅱ/JPI-07Ⅱ

JPI-15Ⅱ/JPI-07Ⅱ

Automated curie point injector

Automated curie point injector

Sample tube Baking Mode
Able to bake-clean mini-PAT (TenaxGR). The baking procedure is pre-programmed and requires a single button operation only.
 
Solvent Purge Mode
Able to remove water and/or solvent trapped in mini-PAT before analysis of VOC.
 
Able to run Py-GC and VOC-GC automatically
 
Replaceable sample tube and needle per analysis
Cross contamination risks are minimized.
 
Curie point heating is an ideal heating method
Large molecules such as n-parrafine up to carbon number 72 can be detected. The entire furnace will not become hot and cause burns. (Patented No4701041 patent application No2011-1422809)

 
   ・Compatible to all major GC models

         

Conventional stationary-pyrolyzers usually block GC injection port and cannot be removed easily. JPI-15II / JPI-07II furnace can slide away to make the GC injection port available for micro syringe injections just by one-touch without removal. 
 
Injection carrier gas is precisely controlled by mass flow controller.
 

 

Sample Tube Baking Mode
Able to bake-clean mini-PAT (Tenax GR). The baking procedure is pre-programmed and requires a single button operation only.
 
Solvent Purge Mode
Able to remove water and/or solvent trapped in mini-PAT before analysis of VOC.
 
Able to run Py-GC/GCMS and VOC-GC/GCMS automatically
 
Replaceable sample tube and needle per analysis
Cross contamination risks are minimized.
 
Curie point heating is an ideal heating method
Large molecules such as n-parrafine up to carbon number 72 can be detected. The entire furnace will not become hot and cause burns. (Patented No4701041 patent application No2011-1422809)
 
Conventional stationary-pyrolyzers usually block GC injection port and cannot be removed easily. JPI-15II / JPI-07II furnace can slide away to make the GC injection port available for micro syringe injections just by one-touch without removal. 
 
Injection carrier gas is precisely controlled by mass flow controller.
 
Compatible to all major GC models.

Specifications

Specifications

Device unit  
 Number of samples JPI-15II : 15 samples
JPI-07II : 7 samples
 Heating method Curie point heating
 Drive system of
 sample tube supply
Rotory drive and vertical motion
 Preheating of sample tube 250°C
 Transfer line 300°C
(Curie point heating)
 Sample tube Quartz glass (OD 6 mm, Length 36 mm)
 Needle OD 0.63mm, Length 42mm
 Dimensions (mm) and weight JPI-15II : 318(W)×330(H)×318(D), 15Kg
JPI-07II : 250(W)×340(H)×250(D), 6Kg

 
Contoroller unit 
 Radio-frequency output 48W, 600kHz
 Operation Panel 4 inch touch panel
 Pyrolysis time 1sec to 30sec
 Signal output GC start signal
 Injection gas He or N2
Dimensions (mm) 180(W)×390(H)×450(D)mm
 Weight 12Kg
 Power AC85 to 240V, 500VA (Max)

 
 
 
 

Divice unit
Number of samples JPI-15II : 15 samples
JPI-07II : 7 samples
Heating method Curie point heating
Drive system of sample tube supply Rotory drive and vertical motion
Preheating of sample tube 250°C
Transer Line 300°C
(Curie point heating)
Sample tube Quartz glass (OD 6 mm, Length 36 mm) 
Needle OD 0.63mm, Length 42mm
Demensions ・JPI-15
318(W) x 330(H) x 318(D) mm
・JPI-07
250(W) x 340(H) x 250(D) mm
Weight JPI-15 : 15 Kg
JPI-07 : 6 Kg

 
Contoroller unit 
Radio-frequency output 48W, 600kHz
Operation panel 4 inch touch panel
Pyrolysis time 1sec  to 99.9sec
Output signal Start signal
Injection gas He or N2
Demension 180(W) x 390(H) x 450(D) mm
Weight 12 Kg
Power AC85 to 240V, 500VA (Max)